Validation of the A&D BP UA-1200BLE device for home blood pressure measurement according to the ISO 81060-2:2013 standard

Buy Article:

$57.00 + tax (Refund Policy)

Objective

The aim of this study was to validate the UA-1200BLE device using the ISO 81060-2:2013 standard.

Materials and methods

Individuals aged more than 12 years were recruited from among the outpatients and volunteers of Jichi Medical University (Tochigi, Japan). Eighty-five patients were included in the final analyses. Validation was performed using the same-arm sequence protocol as per the ISO 81060-2:2013 standard.

Results

The device performed well against the standard; the mean and SD values for the differences between device and observed systolic and diastolic blood pressure values fulfilled both criterion 1 and criterion 2 of the standard (criterion 1: −2.62±6.89 and −0.90±6.63 mmHg, respectively; criterion 2: −2.65±5.36 and −0.91±6.12 mmHg, respectively). The Bland–Altman plots did not show any systematic variation in the error.

Conclusion

The UA-1200BLE device had a high level of accuracy and fulfilled all ISO 81060-2:2013 standard requirements in adults with a wide range of arm circumference measurements. It is therefore suitable as a wireless blood pressure monitoring tool in patients with hypertension.

Keywords: blood pressure; device validation; home blood pressure monitoring; self-blood pressure monitoring; wireless monitoring

Document Type: Research Article

Affiliations: Department of Medicine, Division of Cardiovascular Medicine, Jichi Medical University School of Medicine, Tochigi, Japan

Publication date: 01 December 2018