Design and performance evaluation of a 3-axis force sensor for MIS palpation

Buy Article:

$38.68 + tax (Refund Policy)

Purpose

– This paper aims to develop a novel miniature 3-axis force sensor which can detect the interaction forces during tissue palpation in MIS (minimally invasive surgery). MIS offers many significant merits compared with traditional open surgery, the wound to the patients and the postoperative pains are alleviated and reduced dramatically. However, the inherent drawback due to lack of force feedback still exists while conducting some operation procedures. For example, tissue palpation performed easily during open surgery could not be realized in an MIS manner.

Design/methodology/approach

– The force sensor is based on the resistive-based sensing method that utilizes strain gauges to measure the strain when the external loads are acting on the tip of the sensor. A novel flexible tripod structure with bending and compression deformations is designed to discriminate the magnitudes and directions of the three orthogonal force components. A linear characteristic matrix is derived to disclose the relationship between the sensitivity and the geometric parameters of the structure, and a straightforward geometric parameterized optimization method considering the sensitivity isotropy is proposed to provide the sensor structure with high sensitivity and adequate stiffness.

Findings

– The sensor prototype can perform force measurement with sensing ranges of ± 3.0 N in axial direction and ± 1.5 N in radial direction, and the resolutions are 5 per cent and 1 per cent, respectively. It is concluded that this force sensor is compatible with MIS instruments and the ex-vivo experiment shows that the sensor can be used to perform tissue palpation during MIS procedures.

Originality/value

– This paper is intended to address the significant role of force sensing and force feedback during MIS operations, and presents a new application of the resistive-based sensing method in MIS. A tripod structure is designed and a straightforward optimization method considering the sensitivity isotropy of the sensor is proposed to determine geometric parameters suited for the given external loads.
No Reference information available - sign in for access.
No Citation information available - sign in for access.
No Supplementary Data.
No Article Media
No Metrics

Keywords: Force feedback; Force sensor; MIS; Tissue palpation; Tripod structure

Document Type: Research Article

Affiliations: 1: State Key Laboratory of Robotics and Systems, Harbin Institute of Technology, Harbin, China 2: School of Life Science and Technology, Harbin Institute of Technology, Harbin, China

Publication date: March 16, 2015

  • Access Key
  • Free content
  • Partial Free content
  • New content
  • Open access content
  • Partial Open access content
  • Subscribed content
  • Partial Subscribed content
  • Free trial content
Cookie Policy
X
Cookie Policy
Ingenta Connect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more