Assessment of accuracy, reliability and force measurement errors for a tethered swimming apparatus
Researchers frequently use purpose-built devices to calculate propulsive forces during tethered swimming. Although such devices are subject to force measurement errors, no specific methods have been suggested in the swimming literature for the estimation of these errors. The purpose
of this study was to access the accuracy and reliability of a portable device that is used to measure propulsive forces in tethered swimming, and to estimate the errors caused by hysteresis, sensitivity and linearity. The force values recorded during a maximum front crawl test of an international
level swimmer were used to provide an example of the extent to which measurement errors could affect the values collected during a tethered swimming study. The calculations revealed small and acceptable errors. When summing the errors from all sources, the total errors affecting the minimum,
average and peak forces recorded during the case study were 1.15%, 0.94% and 0.86% respectively. It is recommended that investigators always calculate and report such errors for tethered swimming studies. The methods used in the present study are reasonably simple and not time-consuming, and
could be used when assessing errors for similar tethered swimming devices.
Keywords: BIOMECHANICS; FRONT CRAWL; KINETICS; PROPULSIVE FORCES
Document Type: Research Article
Publication date: 01 December 2011
- Editorial Board
- Information for Authors
- Subscribe to this Title
- Terms & Conditions
- Ingenta Connect is not responsible for the content or availability of external websites
- Access Key
- Free content
- Partial Free content
- New content
- Open access content
- Partial Open access content
- Subscribed content
- Partial Subscribed content
- Free trial content