Open Access An Evaluation of Mass-Weighted Size Distribution Measurements with the Model 3320 Aerodynamic Particle Sizer

Download Article:
 Download
(PDF)
 

Abstract:

The ability of the Model 3320 aerodynamic particle sizer (APS) to make accurate mass-weighted size distribution measurements was investigated. Significant errors were observed in APS size distribution measurements with measured mass median aerodynamic diameters (MMADs) as much as 17 times higher than from cascade impactor measurements. Analysis of APS correlated time-of-flight and light scattering data indicated that the MMAD distortions were due to a few anomalous large particle measurements (~0.1% of the total measurements) with surprisingly low scattered light. Computational fluid dynamics modeling indicated that these anomalous measurements were due to particles that deviated from the intended aerosol pathway and recirculated through the APS measurement volume at low velocities leading to erroneous large particle measurements. A technique for removing erroneous measurements based on correlated aerodynamic diameter and light scattering values is presented. When this technique was used, APS and cascade impactor size distribution measurements agreed well.

Document Type: Research Article

DOI: https://doi.org/10.1080/02786820290092087

Publication date: 2002-07-01

More about this publication?
  • Access Key
  • Free ContentFree content
  • Partial Free ContentPartial Free content
  • New ContentNew content
  • Open Access ContentOpen access content
  • Partial Open Access ContentPartial Open access content
  • Subscribed ContentSubscribed content
  • Partial Subscribed ContentPartial Subscribed content
  • Free Trial ContentFree trial content
Cookie Policy
X
Cookie Policy
Ingenta Connect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more