Skip to main content

Studies on the application of a low-voltage peak to the postsurgical follow-up CT scan in abdominal cancer patients in order to reduce the exposure of patients to radiation

Buy Article:

$71.00 + tax (Refund Policy)

This study examined the radiation dose, computed tomography (CT) number, contrast and image quality of patients requiring periodic follow-up abdominal CT examinations at various tube voltages. The subjects were divided into two groups. One group consisted of patients who underwent a clinical analysis and the other group was a phantom one. Somatom Sensation 16 (Siemens, Erlangen, Germany) was used. Twenty patients who underwent a periodic follow-up examination by CT were selected randomly. The tube current was fixed to 150 mA, and the tube voltage was adjusted according to the appropriate value of each examination. The computed tomography dose index (CTDI) values were measured. The CT number of each organ was measured by setting up a 1 cm diameter return on investment (ROI) in the abdominal organs at the same height of the first lumbar vertebra using images of the arterial phase. Two radiologists in consensus graded the quality of the abdominal images into three groups. An abdomen-shaped acrylic phantom was used in the phantom study. An ion chamber was inserted into the holes located at the center and periphery of the phantom, where the radiation dose was automatically displayed on the reader. Tube voltages of 80, 100, 120 and 140 kVp were applied to the phantom (diluted contrast medium with water at 1:10 ratio) and the phantom was scanned. The CT number was measured from a 1 cm diameter ROI at the center of the image. The CTDI value decreased by 36% at 100 kVp (7.50 mGy) compared with that at 120 kVp (11.70 mGy). According to the radiologists’ evaluation, there were 17 equivalent, 3 acceptable and 0 unacceptable levels in the group of 20 subjects. The radiation dose in the phantom study decreased with increasing tube voltages from 80 to 140 kVp. The peripheral and central doses decreased by 38% and 41%, respectively. The CT numbers at 80, 100, 120 and 140 kVp were 1365.9±4.4, 1046.1±3.7, 862.8±3.2 and 737.5±3.0 HU, respectively. In conclusion, in a follow-up observation for the detection of a recurrence or metastasis after surgery or patients with chronic abdominal diseases, the exposure doses can be reduced using a low-voltage peak CT examination without greatly changing the image quality.

Keywords: CT number; CTDI; radiation dose; tube voltage

Document Type: Research Article

Affiliations: 1: Department of Radiological Science,Gyeongsan University College, Gyeongsan, Republic of Korea 2: Department of Computer Science and Engineering,Soonchunhyang University, Asan, Republic of Korea 3: Department of Radiology,Soonchunhyang University Hospital, Bucheon, Republic of Korea 4: Department of Radiological Technology,Gwangju Health College University, Gwang Ju, Republic of Korea 5: Department of Nuclear Engineering,Chosun University, Dong-gu, Republic of Korea

Publication date: 01 October 2012

  • Access Key
  • Free content
  • Partial Free content
  • New content
  • Open access content
  • Partial Open access content
  • Subscribed content
  • Partial Subscribed content
  • Free trial content