Subsurface and Transcutaneous Raman Spectroscopy and Mapping Using Concentric Illumination Rings and Collection with a Circular Fiber-Optic Array

$29.00 plus tax (Refund Policy)

Buy Article:

Abstract:

Different spatial separations between an illumination ring and a bundle of 50 collection fibers focused to collect light in the center of the ring were used to investigate the recovery of subsurface Raman spectra. The depth of Raman signal recovery and the preservation of spatial information in the recovered signal were investigated using polymer blocks stacked in different geometries. The illumination rings were then combined into a single data set to increase variation in the signal. Multivariate data analysis was used to recover the Raman spectra of the subsurface component. The Raman spectrum of a Delrin target was recoverable at depths up to 22.6 mm of overlying Teflon. Spatial information was lost at approximately 6.5 mm below the Teflon surface. The same protocols were used to recover canine bone spectra transcutaneously at depths up to 5 mm below the skin's surface. The recovered bone spectra were validated by exposed bone measurements.

Keywords: FIBER-OPTIC PROBE; MAPPING; PHOTON MIGRATION; RAMAN SPECTROSCOPY; TOMOGRAPHY; TRANSCUTANEOUS

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/000370207781393307

Affiliations: 1: University of Michigan, Department of Chemistry, Ann Arbor, Michigan 48109-1055 2: University of Michigan, Department of Orthopaedic Surgery, Orthopaedic Research Laboratories, Ann Arbor, Michigan 48109-2200

Publication date: July 1, 2007

More about this publication?
Related content

Tools

Favourites

Share Content

Access Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content
Cookie Policy
X
Cookie Policy
ingentaconnect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more