Spectral Simulation Methodology for Calibration Transfer of Near-Infrared Spectra

Authors: Sulub, Yusuf1; Small, Gary W.2

Source: Applied Spectroscopy, Volume 61, Issue 4, Pages 70A-84A and 349-457 (April 2007) , pp. 406-413(8)

Publisher: Society for Applied Spectroscopy

Buy & download fulltext article:

OR

Price: $29.00 plus tax (Refund Policy)

Abstract:

A spectrum simulation method is described for use in the development and transfer of multivariate calibration models from near-infrared spectra. By use of previously measured molar absorptivities and solvent displacement factors, synthetic calibration spectra are computed using only background spectra collected with the spectrometer for which a calibration model is desired. The resulting synthetic calibration set is used with partial least squares regression to form the calibration model. This methodology is demonstrated for use in the analysis of physiological levels of glucose (1–30 mM) in an aqueous matrix containing variable levels of alanine, ascorbate, lactate, urea, and triacetin. Experimentally measured data from two different Fourier transform spectrometers with different noise levels and stabilities are used to evaluate the simulation method. With the more stable instrument (A), well-performing calibration models are obtained, producing a standard error of prediction (SEP) of 0.70 mM. With the less stable instrument (B), the calibration based solely on synthetic spectra is less successful, producing an SEP value of 1.58 mM. For cases in which the synthetic spectra do not describe enough spectral variance, an augmentation protocol is evaluated in which the synthetic calibration spectra are augmented with the spectra of a small number of experimentally measured calibration samples. For instruments A and B, respectively, augmentation with measured spectra of nine samples lowers the SEP values to 0.64 and 0.85 mM.

Keywords: CALIBRATION TRANSFER; GLUCOSE; NEAR-INFRARED SPECTROSCOPY; PARTIAL LEAST SQUARES; PLS; SIMULATION

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/000370207780466280

Affiliations: 1: Department of Chemistry and Optical Science and Technology Center, University of Iowa, Iowa City, Iowa 52242; Novartis Pharmaceuticals Corp., One Health Plaza, Bldg. 401/B244B, East Hanover, NJ 07936-1080 2: Department of Chemistry and Optical Science and Technology Center, University of Iowa, Iowa City, Iowa 52242

Publication date: April 1, 2007

More about this publication?
Related content

Tools

Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content

Text size:

A | A | A | A
Share this item with others: These icons link to social bookmarking sites where readers can share and discover new web pages. print icon Print this page