@article {Cabalo:2002:0003-7028:1345, author = "Cabalo, Jerry C. and Schmidt, John and Wendt, Jost O.L. and Scheeline, Alexander", title = "Spectrometric System for Characterizing Drop and Powder Trajectories and Chemistry in Reactive Flows", journal = "Applied Spectroscopy", volume = "56", number = "10", year = "2002", abstract = "An instrument for emission, fluorescence, and scattering diagnostics of a 22 kW gas/air thermal reactor is described. Z-pair parabolic mirrors provide diffraction-limited imaging in field center, with worst-case blur over a 12.8 mm field of view having a radius of 0.044 mm rms. Either full-field filtered or transverse line dispersed viewing can occur simultaneously. Dispersion is achieved using a 2-m spectrograph f-matched to the feed optics. Examples of data show non-axisymmetric flow in a nominally laminar flow system, drop shapes in situ, and Mie scattering from soot.", pages = "1345-1353", url = "http://www.ingentaconnect.com/content/sas/sas/2002/00000056/00000010/art00015", doi = "doi:10.1366/000370202760355073", keyword = "FLAME DIAGNOSTICS, PLANAR LASER-INDUCED FLUORESCENCE, SPECTROSCOPY IN HARSH ENVIRONMENTS, IMAGING, DROP VAPORIZATION" }