Influence of Solution-Deposited Anions on Glow Discharge Relative Ion Yields

$29.00 plus tax (Refund Policy)

Buy Article:

Abstract:

When standard addition methods were employed for quantification in glow discharge mass spectrometry (GDMS), relative ion yields for some, but not all, elements were observed to be greater when deposited as a solution residue than as a corresponding solid of the same composition. The origin of these effects was investigated and determined to result from the deposition of anions present in the solution standards. The enhancements could be duplicated from a solid by pretreatment with a solution containing the anion. The mechanism by which the enhancements occur is not well understood, but we believe it involves a form of reactive etching followed by collisional dissociation of the sputtered molecular species into metal ions. Preliminary research suggests that it may be possible to compensate for changes in relative ion yield with solution treatment by employing an appropriate internal reference standard, or to match closely the matrices of the sample and standard in such a manner that quantification by standard addition is possible.

Keywords: Glow discharge; Mass spectrometry; Relative ion yields; Standard addition

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/0003702944028083

Affiliations: 1: Department of Chemistry, Franklin and Marshall College, Box 3003, Lancaster, Pennsylvania 17604 2: Chemical and Analytical Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6375

Publication date: November 1, 1994

More about this publication?
Related content

Tools

Favourites

Share Content

Access Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content
Cookie Policy
X
Cookie Policy
ingentaconnect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more