Measurement of Hemoglobin in Unlysed Blood by Near-Infrared Spectroscopy

Buy Article:

$29.00 plus tax (Refund Policy)

Abstract:

Visible and near-infrared transmittance (T) spectra of unlysed blood samples were obtained with an NIRSystems Model 6500 spectrophotometer modified for an open cell and a vertical light path. Without rigid control of the pathlength and temperature, we were able to measure hemoglobin content within a standard error of 0.43 g/dL using a single-term second-derivative ratio of log(1/T) data at 1740 and 1346 nm. Calibration was done on a set of 104 samples (two spectra of blood from 52 patients) having hemoglobin levels of 6.1 to 19.2 g/dL. Validation was done on an independent set of 56 samples (two spectra of blood from 28 patients) having hemoglobin levels of 7.2 to 19.0 g/dL. The reproducibility of the measurement, tested by computing the coefficient of variability of the 28 duplicated results, was 0.63% (<0.1 g/dL). The results demonstrate a rapid simple diffuse transmittance measurement of hemoglobin in unlysed blood.

Keywords: Concentration; Hemoglobin; Near-infrared

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/000370294775269036

Affiliations: 1: Department of Pathology, The Valley Hospital, 223 North Van Dien Avenue, Ridgewood, New Jersey 07450 2: Consultant, 11204 Montgomery Road, Beltsville, Maryland 20705 3: Department of Biostatistics, Henry M. Jackson Foundation, 1401 Rockville Pike, Suite 600, Rockville, Maryland 20852

Publication date: April 1, 1994

More about this publication?
sas/sas/1994/00000048/00000004/art00010
dcterms_title,dcterms_description,pub_keyword
6
5
20
40
5

Access Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content
Cookie Policy
X
Cookie Policy
ingentaconnect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more