The Use of a Near-Infrared Array Detector for Raman Spectroscopy Beyond One Micron

Buy Article:

$29.00 plus tax (Refund Policy)

Abstract:

The performance characteristics of a multielement Ge array detector have been measured. The effects of read noise and dark signal on signal integration have been evaluated. The performance of this array for Raman spectroscopy with excitation at one micron has been compared with FT-Raman spectroscopy. It is an excellent detector for narrow-spectral-bandwidth Raman studies, and the sensitivity compares favorably with that for FT-Raman techniques.

Keywords: Instrumentation; Raman spectroscopy

Document Type: Rapid Communication

DOI: http://dx.doi.org/10.1366/0003702914336156

Affiliations: 1: Central Research Department, E. I. Dupont, Wilmington, Delaware 19880-0328 2: Princeton Instruments, Princeton, New Jersey 08540