Skip to main content

Identification of Several Wavelengths Useful for ICP that Are Incorrectly Listed in the Literature

Buy Article:

$29.00 plus tax (Refund Policy)

Abstract:

A previously characterized sequential scanning ICP spectrometer with nominal resolution of 0.009 nm, FWHM, for a 3600-line/mm grating and 0.018 nm for a 1800-line/mm grating is used to examine wavelength positions. Five atomic lines believed to have been reported in error in the literature are identified. Alternative values for each wavelength are suggested.

Keywords: ICP emission spectroscopy; Wavelengths

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/0003702914337227

Affiliations: The Perkin-Elmer Corporation, 761 Main Avenue, Norwalk, Connecticut 06859-0219

Publication date: March 1, 1991