Abstract:

X-ray fluorescence results of the analysis of titanium in NIST SRM 1165 and 1264 low-alloy steels were found to be significantly different from the certified values. These differences appear to be unique to the XRF results since results from other methods such as emission spectroscopy did not show any discrepancies. The microhomogeneity of selected samples was examined by electron probe microanalysis and revealed the presence of titanium inclusions containing varying amounts of zirconium, niobium, and tantalum. Differences in metallurgical uniformity are believed to be the major cause of apparent XRF bias in these SRMs. XRF analysis of the new 1760-series alloys indicates that these are more uniform and more accurate for purposes of calibration than the 1160 series and 1260 series for elements such as titanium.

Keywords: Electron probe microanalysis; Inclusions; Low-alloy steels; Titanium; X-ray fluorescence

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/0003702914337614

Affiliations: National Institute of Standards and Technology, Center for Analytical Chemistry, Gaithersburg, Maryland 20899

Publication date: February 1, 1991

More about this publication?
Related content

Tools

Favourites

Share Content

Access Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content
Cookie Policy
X
Cookie Policy
ingentaconnect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more