Performance of an FT-IR with a Cube-Corner Interferometer

$29.00 plus tax (Refund Policy)

Buy Article:

Abstract:

The stability and performance of a Fourier transform infrared spectrometer with a cube-corner interferometer is investigated. Evaluation of instrument performance is based on results from interferogram subtraction experiments, 100% line interpretations, and comparison with theory. It is demonstrated that cube-corner retroreflectors compensate for interferometer moving-mirror tilt, making possible accurate spectroscopic measurements well into the visible region of the electromagnetic spectrum. In addition, cube-corner interferometers have potential for true dual-beam FT-IR measurements.

Keywords: FT-IR; Infrared; Interferometry, infrared region; Optics

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/0003702854248827

Affiliations: Mattson Instruments, 6333 Odana Road, Madison, Wisconsin 53719