Calibrating Raman Spectrometers with Plasma Lines from the Argon Ion Laser

$29.00 plus tax (Refund Policy)

Buy Article:

Abstract:

The majority of the plasma lines of the argon ion laser are shown to be transitions of the argon ion. These lines are useful for calibrating Raman spectrometers. A table is given of the plasma lines, including relative intensities, wavelengths (in standard air), wavenumber (in vacuo), and wavenumber shifts relative to the 488.0 and 514.5 nm laser lines. Spectra showing the relative intensities of these lines are also shown.

Keywords: Argon ion laser; Calibration techniques; Raman spectrometer

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/0003702794925237

Affiliations: 1: Laboratory of Chemical Physics, National Institute of Arthritis, Metabolism and Digestive Diseases, National Institutes of Health, Bethesda, Maryland 20205; On sabbatical leave from the Department of Chemistry, Oberlin College, Oberlin, OH 44074 2: Laboratory of Chemical Physics, National Institute of Arthritis, Metabolism and Digestive Diseases, National Institutes of Health, Bethesda, Maryland 20205

Publication date: September 1, 1979

More about this publication?