Optimum Sample Concentration for Resonance Raman Spectra

Buy Article:

$29.00 plus tax (Refund Policy)


In resonance Raman spectroscopy, absorption of incident radiation and of scattered radiation increases with sample concentration. Scattering intensity at the focal point of the laser also increases with sample concentration. An optimum sample concentration is sought where the scattered radiation intensity emerging from the sample cell is at a maximum. Assuming a point-scattering model, Strekas et al. have derived the relationship:

I = I0Jce−kc(r8 + li

I is the observed intensity, I 0 the laser intensity on entry into the sample cell, J is the molar scattering coefficient, c is the molar concentration, s and i are the decadic molar absorptivities at scattering wavelength and the incident wavelength, r and l are the pathlengths of the laser beam and the scattered radiation in the Raman cell, respectively, and k ∞ ln 10 = 2.303. This function has a maximum. Setting the differential with respect to concentration equal to zero gives the optimum concentration:

c opt = [k(r 8 + l i )]−1

If r = l, Eq. (2) reduces to:

c opt = [kr( 8 + l i )]−1

Keywords: Concentration optimization; Resonance Raman; Self-absorption

Document Type: Short Communication


Affiliations: Eastern Regional Research Center, Agricultural Research Service, U.S. Department of Agriculture, Philadelphia, Pennsylvania 19118

Publication date: May 1, 1978

More about this publication?

Access Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content
Cookie Policy
Cookie Policy
ingentaconnect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more