Optimized Sampling in the Gas Chromatography-Infrared Spectroscopy Interface

$29.00 plus tax (Refund Policy)

Buy Article:

Abstract:

The optimum volume for a flow-through gas cell used in the gas chromatography-infrared spectroscopy interface (GC-ir) is shown to be equal to the volume of carrier gas between the half-width points of the GC peak. The optimum length and diameter for lightpipe gas cells for GC fractions separated using ⅛ in. packed columns and support-coated open tubular columns has been calculated for infrared measurements made using a rapid-scanning Fourier transform infrared spectrometer with either a triglycine sulfate or a mercury cadmium telluride detector. A sampling trigger actuated by the inflection points of the GC peak has been shown to be superior to the more usual threshold type of trigger. Expressions for the delay time between actuation of the trigger and initiation of data collection are also derived.

Keywords: Fourier transform infrared spectroscopy; Gas chromatography; Infrared accessories

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/000370277774463562

Affiliations: Department of Chemistry, Ohio University, Athens, Ohio 45701

Publication date: July 1, 1977

More about this publication?
Related content

Tools

Favourites

Share Content

Access Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content
Cookie Policy
X
Cookie Policy
ingentaconnect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more