Skip to main content

The Effect of Temperature on Carbon Furnace Atomic Emission Spectrometry

Buy Article:

$29.00 plus tax (Refund Policy)

Abstract:

The effect of atomization temperature on detection limits attained in carbon furnace atomic emission spectrometry is described and discussed. Standard carbon tubes have been modified, to give a maximum tube temperature 370°K higher, by removing 1 mm of the outside of the tube at its center. The dependence of the emission signal on temperature is shown to have an exponential form, and detection limits for 25 elements are reported in both standard and modified tubes. The potential analytical advantages of a purely electrothermal energy source in emission spectroscopy are outlined.

Keywords: Emission spectroscopy, Instrumentation, emission spectroscopy; Graphite or carbon furnace spectroscopy

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/000370277774463175

Affiliations: Department of Pure and Applied Chemistry, University of Strathclyde, Cathedral Street, Glasgow G1 1XL, Scotland, United Kingdom

Publication date: January 1, 1977

More about this publication?
sas/sas/1977/00000031/00000001/art00003