Ultratrace Analyses by Optical Emission Spectroscopy: The Stray Light Problem

Buy Article:

$29.00 plus tax (Refund Policy)

Abstract:

When atomic emission spectroscopy is employed for the determination of trace elements at concentrations near the detection limit, the spectral background will normally be a large fraction of the total measured signal. Precise background corrections are therefore required if accurate analyses are to be achieved. Experimental evidence is presented to document the fact that stray light may produce substantial shifts in the background signal measured with some spectrometers when the total composition of the sample changes. The data presented have been obtained with the inductively coupled plasma as an excitation source; similar effects can be expected with other excitation sources as well. Examples of various forms of stray light originating from grating defects (ghosts, near and far scatter) and defects in the design of spectrometers are presented. Various methods for the reduction, elimination or correction of stray light effects are also discussed.

Keywords: Emission spectroscopy; Matrix effects; Stray light

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/000370276774456994

Affiliations: Ames Laboratory ERDA and Department of Chemistry, Iowa State University, Ames, Iowa 50011

Publication date: July 1, 1976

More about this publication?
sas/sas/1976/00000030/00000004/art00002
dcterms_title,dcterms_description,pub_keyword
6
5
20
40
5

Access Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content
Cookie Policy
X
Cookie Policy
ingentaconnect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more