Resolution and Frequency Calibration Performance of Beckman IR-9 Infrared Spectrophotometer

$29.00 plus tax (Refund Policy)

Buy Article:

Abstract:

Modifications affecting the resolution and frequency calibration capability of the Beckman IR-9 infrared spectrophotometer are described and discussed. Under favorable conditions these modifications enable the instrument to be calibrated with an absolute accuracy of 0.03 cm−1 or better; resolution is also improved so that spectral slitwidths may be reduced by approximately √2 over normal specifications. These results are illustrated with vibration-rotation spectra from NH3, CH4, and HBr.

Keywords: Infrared spectra; Infrared, general; Instrumentation, infrared

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/000370274774331994

Affiliations: Centre College, Danville, Kentucky 40422