Simple Helium Enclosure for Increasing Diffraction-Line Detectability with a G.E. XRD-5 Diffractometer

Buy Article:

$29.00 plus tax (Refund Policy)

Abstract:

Various means may be used to improve the detectability of an x-ray diffractometer, such as using pulse-height selection, etc. One method of increasing detectability is that of using a helium path instead of an air path for the x-rays. A graph comparing the detectability of diffraction-line intensities in a helium path to that in an air path for various wavelengths is given in Fig. 1. Chromium K alpha (CrKα) radiation is used for greater dispersion of large d spacings such as are encountered in many organic compounds. The graph in Fig. 1. indicates that the detectability for CrKα is increased approximately 2½ times by using a helium path instead of an air path. A description follows for a simple method of constructing a helium enclosure for the G.E. XRD-5 diffractometer.

Document Type: Short Communication

DOI: http://dx.doi.org/10.1366/000370268774383525

Affiliations: X-Ray Crystallography Laboratory, UNIVAC, St. Paul, Minnesota

Publication date: May 1, 1968