Skip to main content

A Digital Recorder for Mass Spectra

Buy Article:

$29.00 plus tax (Refund Policy)

Abstract:

As a mass spectrum is scanned, the maximum value of each peak is stored in a capacitor. Then the difference between the stored voltage and the lesser voltage seen after the scan has proceeded down the side of the peak operates a peak detector, which in turn initiates measurement of the stored peak height by a digital voltmeter. At the same time the mass is determined by measuring the ion accelerating voltage with a modified self balancing potentiometer equipped with a shaft position digitizer.

Document Type: Research Article

DOI: http://dx.doi.org/10.1366/000370258774615573

Affiliations: Spectroscopy Laboratory, The Dow Chemical Company, Midland, Mich.