Skip to main content

Noise exposure is increased with neonatal helmet CPAP in comparison with conventional nasal CPAP

Buy Article:

$51.00 plus tax (Refund Policy)

Abstract:

Background:

In adults, noninvasive ventilation via a helmet is associated with significantly greater noise than nasal and facial masks. We hypothesized that noise exposure could be increased with neonatal helmet continuous positive airway pressure (CPAP) in comparison with conventional nasal CPAP (nCPAP). Our primary objective was to compare the noise intensity produced by a neonatal helmet CPAP and a conventional nCPAP system. Furthermore, we aimed to evaluate the effect of the gas flow rate and the presence of the humidifier and the filter on noise levels during neonatal helmet CPAP treatment. Methods:

In this bench study, noise intensity was measured in the following settings: helmet CPAP, nCPAP, incubator and the neonatal intensive care unit. In helmet CPAP, noise measurements were performed at different gas flow rates (8, 10 and 12 l/min), while in nCPAP, the flow rate was 8 l/min. For both CPAP systems, the level of pressure was maintained constant at 5 cmH2O. Results:

During neonatal helmet CPAP, the median (interquartile range) noise levels were significantly higher than those during nCPAP: 70.0 dB (69.9–70.4) vs. 62.7 dB (62.5–63.0); P<0.001. In the helmet CPAP, the noise intensities changed with increasing flow rate and with the presence of a humidifier or a filter. Conclusions:

Noise intensities generated by the neonatal helmet CPAP were significantly higher than those registered while using a conventional nCPAP system. In the helmet, the noise intensity depends on the gas flow rate, and the presence of a humidifier and a filter in the system.

Document Type: Research Article

DOI: http://dx.doi.org/10.1111/j.1399-6576.2010.02356.x

Affiliations: 1: Pediatric 2: Information Engineering, Medical School, University of Padua, Azienda Ospedaliera di Padova, Padua, Italy

Publication date: January 1, 2011

Access Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Partial Open Access Content
Partial Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content
Cookie Policy
X
Cookie Policy
Ingenta Connect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more