Expiratory tidal volume displayed on Bird 8400STi can exceed the preset tidal volume due to cardiogenic oscillation: a lung model study

Authors: Imanaka, Hideaki1; Takeuchi, Muneyuki1; Tachibana, Kazuya1; Nishimura, Masaji2

Source: Journal of Anesthesia, Volume 18, Number 4, November 2004 , pp. 313-315(3)

Publisher: Springer

Buy & download fulltext article:

OR

Price: $47.00 plus tax (Refund Policy)

Abstract:

We noticed that monitored tidal volumes often exceeded preset tidal volumes in patients with large cardiogenic oscillation. To investigate whether triggering modes affect this discrepancy, we simulated cardiogenic oscillation of 90 breaths/min in a lung model, which was ventilated with a Bird 8400STi ventilator (Bird, Palm Springs, CA, USA). The magnitude of cardiogenic oscillation was defined as peak expiratory flow fluctuation at the lung model. Two respiratory rates (5 and 10 breaths/min) and two triggering modes (flow-triggering and pressure-triggering) were applied, while tidal volume was set at 500 ml. We recorded tidal volume on a ventilator monitor and calculated the discrepancy from the set tidal volume. We also measured fluctuation in flow and airway opening pressure created by cardiogenic oscillation. During flow-triggering, larger flow fluctuation and smaller airway pressure fluctuation were observed compared with during pressure-triggering. During flow-triggering, the discrepancy between monitored tidal volume and set value ranged from 0 to +327 ml at 5 breaths/min, and from 0 to +105 ml at 10 breaths/min. There was a linear correlation between the magnitude of cardiogenic oscillation and the overestimation of tidal volume. By contrast, during pressure-triggering, the discrepancy was small. In conclusion, tidal volume is overestimated during flow-triggering but not during pressure-triggering when cardiogenic oscillation is large.

Keywords: Cardiogenic oscillation; Tidal volume; Lung model; Flow-triggering; Pressure-triggering

Document Type: Research article

DOI: http://dx.doi.org/10.1007/s00540-004-0257-1

Affiliations: 1: Surgical Intensive Care Unit, National Cardiovascular Center, 5-7-1, Fujishiro-dai, Suita, 565-8565, Japan, 2: Department of Emergency and Critical Care Medicine, Tokushima University School of Medicine, Tokushima, Japan,

Publication date: 2004-11-01

Related content

Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content

Text size:

A | A | A | A
Share this item with others: These icons link to social bookmarking sites where readers can share and discover new web pages. print icon Print this page