A Novel, Versatile Valved Holding Chamber for Delivering Inhaled Medications to Neonates and Small Children: Laboratory Simulation of Delivery Options

Authors: DiBlasi, Robert M; Coppolo, Dominic P; Nagel, Mark W; Doyle, Cathy C; Avvakoumova, Valentina I; Ali, Rubina S; Mitchell, Jolyon P

Source: Respiratory Care, Volume 55, Number 4, April 2010 , pp. 419-426(8)

Publisher: The Journal Respiratory Care Company

Buy & download fulltext article:

The full text article is not available for purchase.

The publisher only permits individual articles to be downloaded by subscribers.

Abstract:

BACKGROUND: Delivery of bronchodilator to infants and small children from a pressurized metered-dose inhaler with valved holding chamber (pMDI-VHC) is limited by airway narrowness, short respiratory cycle time, and small tidal volume (VT). There is a need for a versatile, efficient VHC, given the variety of treatment modalities. METHODS: We tested the AeroChamber Mini VHC (the internal geometry of which is optimized for aerosol delivery, and which accepts a pMDI canister that has a dose counter) in experiments to determine differences in the delivery of hydrofluoroalkane-propelled albuterol (90 μg/actuation) during: mechanical ventilation via endotracheal tube (ETT); manual resuscitation via ETT; and spontaneous breathing via face mask. We tested 5 units of the AeroChamber Mini VHC per test. We simulated the tidal breathing of a premature neonate (VT 6 mL), a term neonate (VT 20 mL), and a child approximately 2 years old (VT 60 mL). We collected the aerosol on an electret filter and quantitatively assayed for albuterol. RESULTS: The total emitted mass of albuterol per actuation that exited the VHC was marginally greater during spontaneous breathing (12.1 ± 1.8 μg) than during manual resuscitation (10.0 ± 1.1 μg) (P = .046). Albuterol delivery via mechanical ventilation, though comparable with the premature-neonate model (3.3 ± 1.2 μg), the term-neonate model (3.8 ± 2.1 μg), and the 2-y-old-child model (4.2 ± 2.3 μg) (P = .63), was significantly lower than in the spontaneous-breathing and manual-resuscitation models (P < .001). In the neonatal models the total emitted mass was similar with the spontaneous-breathing model (6.0 ± 1.0 μg with the premature-neonate model, 10.5 ± 0.7 μg with the term-neonate model) and the manual-resuscitation model (5.5 ± 0.3 μg premature-neonate model, 10.7 ± 0.9 μg term-neonate model) (P ≥ .46 via one-way analysis of variance). CONCLUSION: The reduced delivery of albuterol during mechanical ventilation (compared to during spontaneous breathing and manual resuscitation via ETT) was probably associated with the saturated atmosphere in the breathing circuit (37°C, relative humidity > 99%), compared to the ambient air (22 ± 1°C, 44 ± 7% relative humidity). The AeroChamber Mini VHC may provide a versatile alternative to VHCs that are designed exclusively for one aerosol treatment modality.

Keywords: neonate; small child; inhaler; aerosol; mechanical ventilation; manual resuscitation; spontaneous breathing; simulation

Document Type: Research article

Publication date: 2010-04-01

More about this publication?

Tools

Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content

Text size:

A | A | A | A
Share this item with others: These icons link to social bookmarking sites where readers can share and discover new web pages. print icon Print this page