Free Content External quality assurance for the laboratory diagnosis of Buruli ulcer disease in Ghana

Download Article:

You have access to the full text article on a website external to Ingenta Connect.

Please click here to view this article on Wiley Online Library.

You may be required to register and activate access on Wiley Online Library before you can obtain the full text. If you have any queries please visit Wiley Online Library


Summary Objective 

To assure the quality of the laboratory diagnosis of Buruli ulcer disease; microscopy and PCR were subjected to external quality assurance (EQA). Methods 

Slides were read by test laboratory staff, followed by blinded re-reading by the controller. Parallel testing of PCR specimens was carried out at the local and external reference laboratory. Slides and PCR specimens with discordant results were subjected to a second reading/testing by the controller to determine the final result. For training purposes, slides and PCR specimens with discrepant results were subsequently re-read/re-tested under supervision at the test laboratory. Results 

Microscopy. First reading: concordance rate 82.9%, discordance rate 17.1%, percentage false negatives 27.1% (sensitivity 72.9%), percentage false positives 10.1% (specificity 89.9%). Second reading: concordance rate 97.9%, discordance rate 2.1%, percentage false negatives 4.2% (sensitivity 95.8%), percentage false positives 0.6% (specificity 99.4%). PCR. First testing: concordance rate 87.9%, discordance rate 12.1%, percentage false negatives 8.2% (sensitivity 91.8%), percentage false positives 19.1% (specificity 80.9%). Second testing: concordance rate 96.2%, discordance rate 3.8%, percentage false negatives 4.7% (sensitivity 95.3%), percentage false-positives 2.1% (specificity 97.9%). Conclusions 

EQA identified deficiencies in the laboratory performance. Corrective action consisted in on-site training and reduced the number of false-negative and false-positive microscopy and PCR results.

Keywords: Buruli ulcer diagnosis; Ghana; PCR; microscopy; quality assurance

Document Type: Research Article


Affiliations: 1:  Bernhard Nocht Institute for Tropical Medicine, Hamburg, Germany 2:  Department of Infectious Diseases and Tropical Medicine, University of Munich, Munich, Germany 3:  Kumasi Centre for Collaborative Research in Tropical Medicine, Kumasi, Ghana

Publication date: November 1, 2006

Access Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Partial Open Access Content
Partial Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content
Cookie Policy
Cookie Policy
Ingenta Connect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more