Investigation of Reproducibility and Error Associated with qPCR Methods using Quantifiler® Duo DNA Quantification Kit

Buy Article:

$51.00 plus tax (Refund Policy)

Abstract:

Abstract: 

Reproducibility of quantitative PCR results is dependent on the generation of consistent calibration curves via accurate volume transfers and instrument performance. A review of 14 standard curves, using two different QuantDuo® standard DNA lots, showed variability of cycle threshold values between assays were larger than those of the Internal PCR Control (IPC). This prompted a set of experiments designed to determine the source of variability. Results showed that error introduced during DNA addition to the plate resulted in little variation. A comparison of seven independent series demonstrated cycle threshold variation between dilutions was larger than the variation expected from repeated samples. Modeling the influence of pipette errors on dilution series accuracy indicated that a more rigorous approach to external calibration curve production is required and showed that improvement in calibration curve stability is expected if the pipette conditions are carefully chosen and/or a single validated curve is utilized as the calibrator.

Keywords: DNA quantification; forensic science; propagation of error; quantitative polymerase chain reaction; real-time polymerase chain reaction; reproducibility

Document Type: Research Article

DOI: http://dx.doi.org/10.1111/j.1556-4029.2010.01460.x

Affiliations: Biomedical Forensic Sciences, Boston University School of Medicine, Boston, MA.

Publication date: September 1, 2010

Access Key

Free Content
Free content
New Content
New content
Open Access Content
Open access content
Subscribed Content
Subscribed content
Free Trial Content
Free trial content
Cookie Policy
X
Cookie Policy
Ingenta Connect website makes use of cookies so as to keep track of data that you have filled in. I am Happy with this Find out more