A comparison of the disposable Ambu® AuraOnce™ Laryngeal Mask with the reusable LMA Classic™ laryngeal mask airway
Authors: Sudhir, G.1; Redfern, D.2; Hall, J. E.3; Wilkes, A. R.4; Cann, C.5
Source: Anaesthesia, Volume 62, Number 7, July 2007 , pp. 719-722(4)
Publisher: Wiley-Blackwell
Abstract:
Summary Single-use supraglottic airway devices are now available and are intended to be comparable with the reusable LMA Classic™ laryngeal mask airway. We performed a randomised cross-over study comparing the Ambu® AuraOnce™ Laryngeal Mask with the LMA Classic. Fifty patients participated in the trial. Success rates for insertion at the first attempt were similar (92% with the Ambu and 84% with the LMA Classic; p = 0.22). The volumes of air required to inflate the cuff to produce a seal were similar, but the cuff pressure was lower for the Ambu Laryngeal Mask (median (IQR [range]) 18 (10-31 [0-100] cmH2O) than the LMA Classic 27 (17-50 [4-90] cmH2O; p = 0.007). Visual analogue scores for ease of insertion were 87 (73-93 [26-97]) mm for the Ambu and 84 (60-89 [18-96]) for the LMA Classic (p = 0.017). Complications were similar in both groups. We suggest that the disposable Ambu Laryngeal Mask is an acceptable alternative to the reusable LMA Classic.Document Type: Research article
DOI: http://dx.doi.org/10.1111/j.1365-2044.2007.05067.x
Affiliations: 1: Locum Consultant Anaesthetist 2: Specialist Registrar 3: Reader 4: Senior Research Fellow 5: Research Co-ordinator, Department of Anaesthetics and Intensive Care Medicine, Wales College of Medicine, Cardiff University, Heath Park, Cardiff CF14 4XN, UK
Publication date: 2007-07-01
- In this: publication
- By this: publisher
- In this Subject: Surgery
- By this author: Sudhir, G. ; Redfern, D. ; Hall, J. E. ; Wilkes, A. R. ; Cann, C.

Shopping cart
Receive new issue alert
Get Permissions