Transmission Shift in Fluorescence Microscopy: Adjustable Interference Filters, Theoretical Considerations and Practical Implications
In interference filters, transmitted light bands can be continuously shifted to shorter wavelengths and the filter's full width of half maximum (FWHM) (synonym: half intensity width/HIW) can be reduced when turned into oblique positions so that the incident axial light no longer runs
perpendicularly to the filter's surface. These modifications of the transmitted light can be described by mathematic functions or particular reference lines empirically evaluated for any filter used. Moreover, potential low intensity leaking rays beyond the filter's principal transmission
band can be shifted or eliminated. The optical effects described can be intensified still more when two interference filters are used as doublet. When the angle of incident light deviates from 90°, the transmitted light is divided into two partial beams being linearly polarized; their
planes of oscillation differ by 90°, their transmission peaks by 3–7%. Thus, the transmitted light can be modulated further with a rotatable polarizer. In fluorescence microscopy, adjustable interference filters can be used for excitation and suppression in the manner described so
that all transmitted light spectra can be modulated in tiny steps and optimally adjusted with regard to the properties of the fluorescent specimen. Several astronomy filters were used for practical evaluations of our method. Because of their particular optical design, these filters could be
used for excitation with low-intensity halogen light so that any photobleeching (fading) or other irritations of native specimens could be avoided. Photomicrographs could be taken from living motile fluorescent specimens with a flash-free from indistinctness caused by movement. Our method
could be successfully carried out with standard laboratory microscopes based on transmitted light as well as with standard industry microscopes fitted with a vertical illuminator for examinations in incident light, and all images could be significantly improved by transmission shift.
Keywords: BANDPASS; CONTRAST; EXCITATION; FLASHLIGHT; FLUORESCENCE MICROSCOPY; IMAGE QUALITY; INTERFERENCE FILTERS; LONGPASS FILTERS; NARROW BAND; PHOTOMICROGRAPHY; SHARPNESS; SUPPRESSION; TRANSMISSION SHIFT
Document Type: Research Article
Publication date: 01 December 2011
- Journal of Advanced Microscopy Research (JAMR) provides a forum for rapid dissemination of important developments in high-resolution microscopy techniques to image, characterize and analyze man-made and natural samples; to study physicochemical phenomena such as abrasion, adhesion, corrosion and friction; to perform micro and nanofabrication, lithography, patterning, micro and nanomanipulation; theory and modeling, as well as their applications in all areas of science, engineering, and medicine.
- Editorial Board
- Information for Authors
- Subscribe to this Title
- Ingenta Connect is not responsible for the content or availability of external websites
- Access Key
- Free content
- Partial Free content
- New content
- Open access content
- Partial Open access content
- Subscribed content
- Partial Subscribed content
- Free trial content